Details, datasheet, quote on part number: TPP 40-105A-J
PartTPP 40-105A-J
CategoryPower => Power Supplies => 8826567
TitleSwitching Power Supplies 40W 2x3 Med open 5V 8A
Description

TRACO Power TPP 40 40W and TPP 65 65W AC/DC power supplies feature a reinforced double I/O isolation system that meets the latest medical safety standards (60601-3 3rd edition, 2 x MOPP). The high efficiency (up to 92% for TPP 40 and up to 93.5% for TPP 65) of these TRACO Power devices allows a high power density for the standard 2.44" x 3.0" packaging format. TPP 40 features a 40C to +70C full load operating temperature range and TPP 65 offers a full load range of 40C to +60C. These devices go up to +85C with 50% load derating (for single output models). TPP 40 and TPP 65 offer active power factor correction and high reliability. Applications include medical devices and demanding safety and space-critical applications.

CompanyTraco Power
DatasheetDownload TPP 40-105A-J datasheet
Specifications 
ProductSwitching Supplies
Output Voltage-Channel 15 VDC
Output Voltage-Channel 2
Number of Outputs1 Output
Output Power40 W
Open Frame/EnclosedOpen Frame
Input Voltage85 VAC to 264 VAC, 120 VDC to 370 VDC
Output Current-Channel 18 A
Output Current-Channel 2
Commercial/MedicalMedical
Isolation Voltage4 kVAC
Mounting StyleChassis
Length89.7 mm
Width60.5 mm
Height34.5 mm
ApprovalsIEC/EN 60601-1, IEC/EN 60950-1, UL 60950-1
SeriesTPP 40
Efficiency90 %
TPP 40-105A-J photo

Others parts numbering
TPP 65-124A-J: Switching Power Supplies 65W 2x3 Med open 24V 2.71A
TPP 65-148A-J: Switching Power Supplies 65W 2x3 Med open 48V 1.36A
TPP 40-112A-J: Switching Power Supplies 40W 2x3 Med open 12V 3.34A
TPP 65-112A-J: Switching Power Supplies 65W 2x3 Med open 12V 5.42A
TPP 65-105A-J: Switching Power Supplies 65W 2x3 Med open 5V 10A
TPP 40-148A-J: Switching Power Supplies 40W 2x3 Med open 48V 0.84A
TPP 40-331M3: Switching Power Supplies 40W 15V 2.67A, 5V 6A, -15V 0.5A
TPP 40-3512: Switching Power Supplies 40W 24V 1.67A, 5V 6A, 12V 0.5A
TPP 65-3512: Switching Power Supplies 65W 24V 2.71A, 5V 8A, 12V 0.6A
TPP 40-124A-J: Switching Power Supplies 40W 2x3 Med open 24V 1.67A
TPP 40-221: Switching Power Supplies 40W 12V 3.34A, 5V 6A
TPP 40-321M2: Switching Power Supplies 40W 12V 3.34A, 5V 6A, -12V 0.5A
TPP 40-251: Switching Power Supplies 40W 24V 1.67A, 5V 6A
TPP 65-251: Switching Power Supplies 65W 24V 2.71A, 5V 8A
TPP 40-231: Switching Power Supplies 40W 15V 2.67A, 5V 6A
TPP 65-231: Switching Power Supplies 65W 15V 4.34A, 5V 8A
TPP 65-221: Switching Power Supplies 65W 12V 5.42A, 5V 8A
TPP 65-321M2: Switching Power Supplies 65W 12V 5.42A, 5V 8A, -12V 0.6A
TPP 65-331M3: Switching Power Supplies 65W 15V 4.34A, 5V 8A, -15V 0.6A

 

Features, Applications

High power density power supply (enclosed / open frame) Certification according to IEC/EN/ES 60601-1 3rd edition for 2xMOPP Low leakage current <75 A rated for BF applications

Risk management process according to ISO 14971 including risk management file Acceptance criteria for electronic assemblies according to IPC-A-610 Level 3

Protection class l and ll Operating to 5000m altitude Ready to meet ErP directive, W no load power consumption 5 year product warranty

The TPP 40 Series of 40 Watt AC/DC power supplies feature a reinforced double I/O isolation system according to latest medical safety standards (60601-1 3rd edition, 2 x MOPP). The earth leakage current is below 75 A what makes the units suitable for BF (body floating) applications. The excellent efficiency to 92% allows a high power density for the standard x 3.0" packaging format. The full load operating temperature range to +70C while it goes to 85C with 50% load derating (for single output models). They come with an active power factor correction and the EMC characteristic is dedicated for applications in industrial and domestic fields. High reliability is provided by use of industrial quality grade components and an excellent thermal management. It makes the products an ideal solution for medical devices and for demanding safety and space critical applications.

Order code Output voltage (Vout 1 adjustable by 10%) Vout 1 5 VDC 12 VDC 15 VDC TPP 40-124A-J TPP 40-148A-J TPP 40-221

Output current max. Vout 3 Vout 0.84 A Vout 2 Vout 3
Total power should not exceed nominal power
Input voltage range Input current at full load

AC range (universal input) 264 VAC DC range 370 VDC at 115 VAC / 230 VAC 1.05 A max. 0.55 A max.

Input frequency 63 Hz Input protection T3.15 A/250 VAC (internal fuse in both line & neutral) Input inrush current at 230 VAC 60 A max. Zero load power consumption 0.15 W max. (acc. ErP directive)

Voltage set accuracy single output:1% multi output: Vout2, Vout3 Regulation - single output Input variation Load variation VDC model: other models: Regulation - multi output Input variation Load variation Vout3: cross regulation max. 0.7% max. 0.5% max. 0.2% max. 0.5% max. 1.5% max. (0.1W to full load: 0.7% max.) 0.7% max. 1.5% max.

Minimum load not required (Vout3 requires 0.5 W over to be stabilized) Temperature coefficient 0.02%/K Hold-up time Vin = 115 VAC 25 ms typ. Start-up time <1s Rise time 20 ms typ. Ripple and noise - single output model (20Mhz Bandwidth) - multi output Vout 1 Vout 2 Vout 3 Current limitation Transiente response 5-15 VDC outputs: 24 VDC output: 48 VDC output: 12 VDC: 15 VDC: 24 VDC 5 VDC: (-)12 VDC: 15 VDC: 75 mVp-p typ. w. cap. 1206 X7R MLCC 75 mVp-p typ. w. cap. 1206 X7R MLCC 150 mVp-p typ. w. cap. 1206 X7R MLCC 120 mVp-p typ. w. cap. 1206 X7R MLCC 150 mVp-p typ. w. cap. 1206 X7R MLCC 240 mVp-p typ. w. cap. 1206 X7R MLCC 100 mVp-p typ. w. cap. 1206 X7R MLCC 120 mVp-p typ. w. cap. 1206 X7R MLCC 150 mVp-p typ. w. cap. 1206 X7R MLCC

Overvoltage protection 140% of nominal Vout single output: at 145% Iout typ. multi output: Pout1 +Pout2

Short circuit protection hiccup mode (automatic recovery) Peak deviation Vout1: 3% max. (25% load step change) Recovery time 600 s typ.

Operating temperature to +85C with derating, see graph on p.3 Output power derating Temperature single output: multi output: Low input voltage 2.67 %/K above at 230 VAC 2.75 %/K above at 115 VAC 1.60 %/K above at 230 VAC 1.67 %/K above at 115 VAC 1.33 %/V below 100 VAC

All specifications valid at nominal input voltage, full load and +25C after warm-up time unless otherwise stated. http://www.tracopower.com Page of 5

Storage temperature to +85C Humidity (non condensing) rel. H max. Altitude during operation 5000 m Switching frequency - single output 5 VDC model: (at 230 VAC) other models: Switching frequency - multi output Vout 1 5 VDC model: (at 230 VAC) other models: Vout 2 Vout 3 Isolation voltage (2 x MOPP insulation) 70 kHz typ. (pulse frequency modulation) 120 kHz typ. (pulse frequency modulation) 70 kHz typ. (pulse frequency modulation) 115 kHz typ. (pulse frequency modulation) 750 kHz typ. (pulse frequency modulation) 510 kHz typ. (pulse frequency modulation)

Input / Output (60 sec.) 4000 VAC Input / Case (60 sec.) 2500 VAC 100 Mohm min.

Leakage current (at 75 A max. Isolation resistance (at 500 VDC) Reliability calculated MTBF at +25C acc. to IEC 3`000`000 h for single output models 1`700`000 h for multi output models Protection class II prepared Electromagnetic compatibility Conducted input RI suppression EN 55022, class B (EMC), emissions Harmonic current emissions IEC EN 61000-3-2, class A Voltage flicker IEC EN 61000-3-3, (class tba.) Radiated input surpression EN 55011, class B IEC EN 60601-1-2 (for medical equipment) Electromagnets compatibility (EMC), immunity Electrostatic discharge ESD RF field immunity Electrical fast transients/burst immunity Surge Conducted RF Magnetic field (only for single output models) IEC EN 60601-1-2 IEC 61000-4-2, 8kV/6kV perf. criteria A IEC 61000-4-3, 20V/m perf. criteria A IEC 2kV perf. criteria A IEC 1kV/ 2kV perf. criteria A IEC 61000-4-6, 20 Vrms perf. criteria A IEC 61000-4-8, 10A/m perf. criteria A

Voltage dip and interruptions according 30%, 500ms perf. criteria A reference: 100 VAC 60%, 100ms perf. criteria 95%, 10ms perf. criteria 95%, 5000ms perf. criteria B Safety standards and certification UL 60950-1, IEC/EN 60950-1, www.ul.com File e188913 IEC/EN 60601-1 3rd edition, ANSI/AAMI ES60601-1:2005(R)2012 Environment Vibration acc. IEC 60068-2-6 3 axis, sine sweep, 1oct/min Shock acc. IEC 60068-2-27 3 axis, 10g half sine, G (3 directions each 3 times) Environmental compliance Reach www.tracopower.com/overview/tpp40 RoHS directive 2011/65/EU Connection screw terminal / pin connector Single output models:on temperature (for horizontal mounting) Power derating depending

Multi derating output models: Power depending on temperature (for horizontal mounting)

All specifications valid at nominal input voltage, full load and +25C after warm-up time unless otherwise stated. http://www.tracopower.com


 

Some Part number from the same manufacture Traco Power
TPP 65-124A-J

TRACO Power TPP 40 40W and TPP 65 65W AC/DC power supplies feature a reinforced double I/O isolation system that meets the latest medical safety standards (60601-3 3rd edition, 2 x MOPP). The high efficiency

TPP 65-148A-J
TPP 40-112A-J
TPP 65-112A-J
TPP 65-105A-J
TPP 40-148A-J
TPP 40-331M3
TPP 40-3512
TPP 65-3512
TPP 40-124A-J
TPP 40-221
TPP 40-321M2
TPP 40-251
TPP 65-251
TPP 40-231
TPP 65-231
TPP 65-221
TPP 65-321M2
TPP 65-331M3
 