Details, datasheet, quote on part number: JHL0312S05
PartJHL0312S05
Category
TitleCONVERTOR DC-DC 3 WATT
Description

XP Power’s JHL and JHM series of encapsulated, isolated, and regulated DC-DC converters are approved for medical applications. Available with single and dual outputs, the 3 W JHL03, the 6 W JHL06, and the 15 W JHM15 are all packaged in industry-standard format cases. Their size is model dependent, ranging from an industry standard DIP-24 for the JHL03 and JHL06 through to 40.6 mm x 25.4 mm x 10.2 mm for the JHM15 series.



CompanyXP Power
DatasheetDownload JHL0312S05 datasheet
  
JHL0312S05 photo

Others parts numbering
JHL0324S05
JHL0312D12
JHL0312S12
JHL0312S15
JHL0612S12: CONVERTOR DC-DC 6 WATT
JHL0612S05
JHL0624S05
JHL0612S15
JHL0612D12
JHM1524S05: DC/DC CONVERTER 15W MEDICAL
JHM1512S12
JHM1512S05
JHM1512D15
JHM1512D05

 

Features, Applications

International Medical Approvals 4000 VAC Reinforced Insulation Meets IEC60601-1, 3rd Edition 2 MOPP Isolation at 250 VAC 2 A Patient Leakage Current DIP24 Package EN55011 Level A With No External Components 3 Year Warranty

1. Input current measured at nominal input voltage. 2. Input current measured at lowest input voltage. 3. Maximum capacitive load is per output. 4. Typical values.

Single -Vin No Pin Trim No Pin +Vout -Vout +Vin
Dual -Vin Common Trim -Vout +Vout Common +Vin

1. All dimensions are in inches (mm) 2. Weight: 0.04 lbs (20 g) approx. 3. Pin diameter: Pin pitch tolerance: (0.35) 5. Case tolerance: 0.02 (0.5)

Characteristic Input Voltage Range Input Current Inrush Current Input Filter Patient Leakage Current Undervoltage Lockout Input Surge >8.8 V. Off >17.5 V. Off 50 Pi type 2 25 Minimum 10 20 Typical Maximum 17 30

Notes & Conditions 12 V nominal 24 V nominal See Models and Ratings table At 30 VDC input

Characteristic Output Voltage Output Voltage Trim Initial Set Accuracy Minimum Load Start Up Delay Start Up Rise Time Line Regulation Load Regulation Cross Regulation Transient Response Ripple & Noise Short Circuit Protection Overload Protection Overvoltage Protection Temperature Coefficient C Minimum 5 Typical Maximum Units ms deviation % pk-pk - 100% load On dual output models with one output set to 50% load and the other varied from to 100% load to 100%) Recovery to within <500 s for a 50% load change at 0.25 A/s rate 20 MHz bandwidth Trip & Restart (hiccup mode), auto recovery Trip & Restart (hiccup mode) Notes & Conditions See Models and Ratings table Via external resistors, see Application Notes V2 of dual output models No minimum load required

Characteristic Efficiency Isolation Patient Leakage Current Input to Output Capacitance Switching Frequency Power Density Mean Time Between Failure Weight Minimum Typical 80 Maximum Units % VAC A pF kHz W/in3 MHrs lb (g) C GB Notes & Conditions See Models and Ratings table For 1 min. Double/reinforced with a working voltage of 250 VAC. Meets 2 x MOPP per 3rd edition IEC60601-1 5000 VAC for ms in accordance with IEC60664-1

Characteristic Operating Temperature Storage Temperature Case Temperature Humidity Cooling Shock Vibration 5 Minimum -20 -40 Typical Maximum Units C %RH Non-condensing Natural convection 3 shocks in each plane, total 18 shocks 11 ms halfsine. Conforms 2 g sweep and endurance at resonance in all 3 planes. Conforms to EN60068-2-6 Notes & Conditions See derating curve

Phenomenon Conducted Radiated Standard EN55011 Test Level A Level A

Phenomenon Immunity ESD Immunity Radiated Immunity EFT/Burst Surges Conducted Immunity Magnetic Fields Safety Approvals Standard EN61000-4-6 EN61000-4-8 Test Level 8 kv Contact, 15 kv Air 10 V/m Vm 30 A/m Criteria As Below 80 MHz - 2.7 GHz plus discrete communication proximity field frequencies Notes & Conditions

Safety Agency CB Report UL EN Safety Standard IEC60601-1 Including Risk Management ANSI/AAMI ES60601-1 3rd Ed. & CSA No.60601-1:2008 EN60601-1 Notes & Conditions Medical

For 12 V output: Trim 12.8 k typical Trim 9.5 k typical For 15 V output: Trim 18 k typical Trim 14.8 k typical

For 12 V output: Trim 5.9 k typical Trim 11.3 k typical For 15 V output: Trim 8.4 k typical Trim 10.4 k typical


 

Some Part number from the same manufacture XP Power
JHL0324S05

XP Power’s JHL and JHM series of encapsulated, isolated, and regulated DC-DC converters are approved for medical applications. Available with single and dual outputs, the 3 W JHL03, the 6 W JHL06,

JHL0312D12
JHL0312S12
JHL0312S15
JHL0612S12
JHL0612S05
JHL0624S05
JHL0612S15
JHL0612D12
JHM1524S05
JHM1512S12
JHM1512S05
JHM1512D15
JHM1512D05
 
0-C     D-L     M-R     S-Z